The seed germination room can realize the function of setting the running time, temperature, humidity, and illumination value of each experimental section. Satisfying all the requirements of seed germination, germination of seeds is no longer a problem. However, when using the intelligent seed germination room, pay attention to the following five points:
1. Although the seed germination box is a new type of experimental equipment that adopts microcomputer control technology and artificially simulates the natural ecological environment, the temperature of each layer in the box is different, and the temperature inside and outside the same layer is also different. Those with larger volumes are There is a certain difference in the temperature of each part of the box, and it should be highly valued when used.
2. From the measurement results, the temperature of the uppermost layer of the light incubator is not the highest, and the temperature in the middle of the box is higher.
3. There are differences between the actual temperature of the seed germination box and the display temperature of the box and the set temperature. The actual temperature in the box is generally smaller than the display temperature and the set temperature of the box. In the actual work, the temperature and humidity recorder is used. Correction.
4. When the number of samples is small and small-scale experiments such as rice germination detection are performed, more accurate seed germination boxes with smaller volumes and more uniform temperatures should be used.
5. When there are a large number of samples and a large volume of experiments, it is necessary to pay special attention to two points when using RTOP-1000D and other light-incubators with large volume and uniform temperature. One is to place a reasonable amount of space in accordance with the temperature requirements of the test subject to ensure that the temperature inside the box can meet the needs of different crops. The second is that it is best not to place samples on the top and bottom layers. If there are a large number of samples and must be used, the samples in the uppermost layer and the lowermost layer should be exchanged with the samples in the higher intermediate temperature region in time, so that the samples in each region can develop synchronously and the prescribed initial/last counts. Tests are completed within days.